TB 9-6625-2294-24
NOTE
Press oscilloscope calibrator CHANNEL pushbutton. Next
press blue soft pushbutton located below CHAN 2 on
oscilloscope calibrator when SELECT CHANNEL is displayed.
(10) Ensure oscilloscope calibrator SOURCE/MEASURE CHAN 2 is connected to TI
CH 2 using a 50 feedthrough termination.
initially for 120 mV. Rotate knob located below EDIT FIELD pushbutton for 6 divisions of
display on TI.
(12) Sweep oscilloscope calibrator frequency to 350 MHz. TI display will be 3.5
divisions in amplitude or greater throughout sweep; if not, perform b below.
(13) Press TRIGGER SOURCE lower pushbutton to CH 3.
(14) Press VERTICAL MODE CH 2 pushbutton off and CH 3 on.
(15) Press CH 3 VOLTS/DIV pushbutton for .1 V.
NOTE
Press oscilloscope calibrator CHANNEL pushbutton. Next
press blue soft pushbutton located below CHAN 3 on
oscilloscope calibrator when SELECT CHANNEL is displayed.
(16) Ensure oscilloscope calibrator SOURCE/MEASURE CHAN 3 is connected to TI
CH 3 using a 50 feedthrough termination.
initially for 600 mV. Rotate knob located below EDIT FIELD pushbutton for 6 divisions of
display on TI.
(18) Sweep oscilloscope calibrator frequency to 350 MHz. TI display will be 3.5
divisions in amplitude or greater throughout sweep; if not, perform b below.
(19) Press TRIGGER SOURCE lower pushbutton to CH 4.
(20) Press VERTICAL MODE CH 3 pushbutton off and CH 4 on.
(21) Press CH 4 VOLTS/DIV pushbutton for .1 V.
NOTE
Press oscilloscope calibrator CHANNEL pushbutton. Next
press blue soft pushbutton located below CHAN 4 on
oscilloscope calibrator when SELECT CHANNEL is displayed.
(22) Ensure oscilloscope calibrator SOURCE/MEASURE CHAN 4 is connected to TI
CH 4 using a 50 feedthrough termination.
initially for 600 mV. Rotate knob located below EDIT FIELD pushbutton for 6 divisions of
display on TI.
(24) Sweep oscilloscope calibrator frequency to 350 MHz. TI display will be 3.5
divisions in amplitude or greater throughout sweep, if not, perform b below.
23