TB 9-6625-2294-35
SECTION
MENU CALIBRATION PROCESS
Preliminary instructions ....................................
Equipment setup.................................................
Power supply .......................................................
CAL 01 horizontal...............................................
CAL 02 vertical ...................................................
CAL 03 triggering ...............................................
CAL 06 vertical transient response...................
Final procedure ...................................................
IDENTIFICATION AND DESCRIPTION
Test Instrument Identification. This bulletin provides instructions for the
was used as the prime data source in compiling these instructions. The equipment being
calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately
6 hours, using the dc and low frequency technique.
Forms, Records, and Reports
a. Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in
which they appear. Report only those adjustments made and designated with (R).
Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.