TB 9-6625-2296-24
(2) Press DATA ENTRY ENTER key to highlight Proper Reading field. Enter
10 using DATA ENTRY keys and press DATA ENTRY ENTER key. Press Ret (F5) key.
(3) Press MODE MTRS (Meter Menu) key; when prompted for backup data press
SHIFT and Y keys (Backup Cal Data ? y/n).
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI DMM V/Ω and COM.
(2) Press key sequence listed in (a) through (f) below:
(a) MODE MTRS ..........................
Meters Menu
(b) AUX (F6) ..................................
Auxiliary Functions Menu
(c) MEMORY RCL .......................
Recall Parameters Menu
(d) FIELD SELECT ..................
Highlight 10. Factory Defaults
(e) DATA ENTRY ENTER ..........
Factory Defaults recall
NOTE
Assure highlighted area indicates YES before proceeding.
(f) DATA ENTRY ENTER ..........
Factory Defaults restored
(3) Press key sequence listed in (a) through (c) below:
(a) MODE MTRS ..........................
Meters Menu
(b) FIELD SELECT ..................
(c) DATA ENTRY ENTER ..........
Select Multimeter
(4) Press Range (F2) key, adjust DATA SCROLL knob to select TI range settings,
press DATA ENTRY ENTER key, and set calibrator for output settings listed in table 6.
TI multimeter indications will be within limits specified in table 6.
Test
Test instrument
instrument
Calibrator output settings
range settings
Amplitude
Min
Max
200 mV
180 mV
1
kHz
169.9 mV
190.1 mV
2.0 V
1.0 V
50
Hz
0.899 V
1.101 V
2.0 V
1.0 V
100
Hz
0.899 V
1.101 V
2.0 V
1.0 V
1
kHz
0.899 V
1.101 V
2.0 V
1.0 V
10
kHz
0.899 V
1.101 V
2.0 V
1.0 V
20
kHz
0.899 V
1.101 V
2.0 V
1.8 V
1
kHz
1.699 V
1.901 V
20
V
18
V
1
kHz
16.99 V
19.01 V
200
V
180
V
100
Hz
169.9
V
190.1
V
V1
2000
470
V
100
Hz
369
V
571
V
1 After
verification press RESET key on calibrator.