TB 9-6625-2296-24
(b) SETUP .....................................
Gen Menu
(c) DATA ENTRY 5 ......................
RF Gen Setup
(d) DATA ENTRY 2 ......................
RF Gen Level
(e) DATA ENTRY 0 ENTER .......
0 dBm
(f) DATA ENTRY 1 ......................
RF Gen Freq
(g) DATA ENTRY 1 M .................
1 MHz
(h) Ret (F5) ....................................
Return
(5) Measuring receiver FM indication will be <45 Hz.
(6) Enter 9 9 . 9 9 9 9 M using TI DATA ENTRY keys.
(7) Measuring receiver FM indication will be <45 Hz.
(8) Enter 5 0 0 M using TI DATA ENTRY keys.
(9) Measuring receiver FM indication will be <140 Hz.
(10) Enter 9 9 9 M using TI DATA ENTRY keys.
(11) Measuring receiver FM indication will be <140 Hz.
(12) Set measuring receiver to measure AM.
(13) Enter 1 M using TI DATA ENTRY keys.
(14) Measuring receiver AM indication will be <0.10%.
(15) Step TI frequency in 100 MHz increments from 100 MHz to 999.9999 MHz using
technique of (13) above. Measuring receiver AM indications will be <0.10%.
b. Adjustments. No further adjustments can be made.
a. Performance Check
(1) Press key sequence listed in (a) through (f) below:
(a) MODE MTRS ..........................
Meters Menu
(b) AUX (F6) ..................................
Auxiliary Functions Menu
(c) MEMORY RCL .......................
Recall Parameters Menu
(d) FIELD SELECT ..................
Highlight 10. Factory Defaults
(e) DATA ENTRY ENTER ..........
Factory Defaults recall
NOTE
Assure highlighted area indicates YES before proceeding.
(f) DATA ENTRY ENTER ........... Factory Defaults restored