TB 9-6625-2296-24
Table 13. Scan Width Accuracy - Continued.
Signal
Measuring receiver indications
Test instrument
Min
Max
Settings
Graticule
initial setting
(kHz)
(kHz)
scan
indication
100 MHz
-225.0
-175.0
50 kHz
2d
100 MHz
175.0
225.0
50 kHz
10th
100 MHz
-450.0
-350.0
100 kHz
2d
100 MHz
350.0
450.0
100 kHz
10th
100 MHz
-900.0
-700.0
200 kHz
2d
100 MHz
700.0
900.0
200 kHz
10th
2001 MHz
-2250.0
-1750.0
500 kHz
2d
200 MHz
1750.0
2250.0
500 kHz
10th
200 MHz
-3500.0
-2500.0
1 MHz
3d
200 MHz
2500.0
3500.0
1 MHz
9th
Set measuring receiver TRACK MODE on.
1
(10) Press FIELD SELECT keys to select Freq and Scan functions and press
DATA SCROLL arrow keys. Adjust DATA SCROLL knob to set Scan and Freq
remaining values. Repeat (3) through (5) and (7) through (9) above for remaining settings
(11) Press key sequence listed in (a) through (g) below:
(a) MODE MTRS ..........................
Meters Menu
(b) AUX (F6) ..................................
Auxiliary Function Menu
(c) MEMORY RCL .......................
Recall
(d) FIELD SELECT ..................
Highlight 10. Factory Defaults
(e) DATA ENTRY ENTER ..........
Factory Defaults recall
NOTE
Assure highlighted area indicates YES before proceeding.
(f) DATA ENTRY ENTER ..........
Factory Defaults restored
(g) SCOPE/ANLZ twice ................
Analyzer Display
(12) Press key sequence listed in (a) through (n) below:
(a) SETUP .....................................
Scope/Analyzer Menu
(b) DATA ENTRY 2 ......................
Analyzer On
(c) DATA ENTRY 4 ......................
Setup Analyzer
(d) DATA ENTRY 2 ......................
(e) DATA ENTRY 1 M .................
1 MHz
(f) DATA ENTRY 4 ......................
Input Atten
(g) DATA ENTRY 3 ......................
40 dB
39