TB 9-6625-2296-24
(a) SETUP .....................................
Scope/Analyzer Menu
(b) DATA ENTRY 4 ......................
Setup Analyzer
(c) DATA ENTRY 2 ......................
Frequency
(d) DATA ENTRY 4 5 0 M ...........
450 MHz
(e) DATA ENTRY 4 ......................
Input Atten
(f) DATA ENTRY 3 ......................
40 dB
(g) Ret (F5) ....................................
Return
(24) Set attenuator to 0 dB, signal generator frequency to 450 MHz and amplitude
for a 0 dB TI indication.
(Normalize) key, and reconnect after allowing time for normalize cycle to complete. If
necessary, adjust signal generator amplitude for a 0 dB TI indication.
(26) Repeat technique of steps (16) through (18) above. TI indications will be within
5 dB of attenuator test report values.
(27) Press key sequence listed in (a) through (g) below:
(a) SETUP .....................................
Scope/Analyzer Menu
(b) DATA ENTRY 4 ......................
Setup Analyzer
(c) DATA ENTRY 2 ......................
Frequency
(d) DATA ENTRY 9 5 0 M ...........
950 MHz
(e) DATA ENTRY 4 ......................
Input Atten
(f) DATA ENTRY 3 ......................
40 dB
(g) Ret (F5) ....................................
Return
(28) Set attenuator to 0 dB, signal generator frequency to 950 MHz and amplitude
for a 0 dB TI indication.
(Normalize) key. Reconnect after allowing time for normalize cycle to complete. If
necessary, adjust signal generator amplitude for a 0 dB TI indication.
(30) Repeat technique of steps (16) through (18) above. TI indications will be within
5 dB of attenuator test report values.
(31) Disconnect signal generator and attenuator from TI.
(32) Press key sequence listed in (a) through (g) below:
Meters Menu
(a) MODE MTRS ..........................
(b) AUX (F6) ..................................
Auxiliary Function Menu
(c) MEMORY RCL .......................
Recall
41