TB 9-6625-2296-24
(2) Press key sequence listed in (a) through (f) below:
(a) MODE MTRS ..........................
Meters Menu
(b) AUX (F6) ..................................
Auxiliary Function Menu
(c) MEMORY RCL .......................
Recall
(d) FIELD SELECT ..................
Highlight 10. Factory Defaults
(e) DATA ENTRY ENTER ..........
Factory Defaults recall
NOTE
Assure that the highlighted area indicates YES before
proceeding to next step.
(f) DATA ENTRY ENTER ..........
Factory Defaults restored
(3) Press key sequence listed in (a) through (i) below:
(a) MODE RF GEN ......................
(b) F1 (Disp) ..................................
Displays Menu
(c) DATA ENTRY 2 (Full Anlz) .
Full Analyzer Display
(d) More (F6) .................................
More
(e) F3 (Level).................................
Select Level Display
(f) DATA ENTRY 0 ENTER .......
0 dB level
(g) More (F6) .................................
More 1
(h) More (F6) .................................
More 2
(i) F5 (Scan)..................................
Select Scan Display
(4) Press DATA SCROLL arrow keys to toggle between 20 kHz and 200 kHz while
adjusting ANALYZER RF A5 ANALZR CENTER (fig. 2) for best centered compromise.
(5) Press key sequence listed in (a) through (h) below:
(a) FIELD SELECT ..................
Highlight SOURCE
(b) DATA SCROLL arrow ...........
Set MOD to FM
(c) DATA ENTRY ENTER ..........
Select FM
(d) FIELD SELECT ..................
Highlight DEVIATION
(e) DATA ENTRY 6 5 K ...............
65 kHz deviation
(f) FIELD SELECT ..................
Highlight AF FREQ
(g) DATA ENTRY 2 0 K ...............
20 kHz mod rate
(h) F5 (Scan)..................................
Select scan
(6) Press DATA SCROLL arrow keys to toggle between 10 kHz and 20 kHz while
adjusting ANALYZER RF A5 DISP ADJUST (fig. 2) for the best dispersion compromise.