TB 9-6625-2296-24
(c) MEMORY RCL .......................
Recall
(d) FIELD SELECT ..................
Highlight 10. Factory Defaults
(e) DATA ENTRY ENTER ..........
Factory Defaults recall
NOTE
Assure highlighted area indicates YES before proceeding.
(f) DATA ENTRY ENTER ..........
Factory Defaults restored
(2) Press MODE RF GEN (MODE RF GEN display), Level (F3) (Level display),
DATA ENTRY 1 2 7 and ENTER keys (-127 dB).
(3) Press key sequence listed in (a) through (f) below:
(a) MODE RCVR ..........................
Receiver Display
(b) SETUP .....................................
Rcvr Menu
(c) DATA ENTRY 1 ......................
Set Rcvr Freq
(d) DATA ENTRY 1 0 0 M ...........
100 MHz
(e) DATA ENTRY 3 ......................
Select Rcvr In T/R
(f) RET (F5) ..................................
Return
(4) Stand TI on end and locate power term, A14 (fig. 6) in lower left hand corner.
(5) Press MODE MTRS (Meters Menu) key and DATA ENTRY 3 key (Pwr Meter
(Pulse/CW)).
(6) Press Range (F1) and DATA SCROLL arrow keys to select 2 W range, then
press DATA ENTRY ENTER key.
(7) Connect a 50 Ω load to TI T/R.
(8) Connect multimeter to A14FL9 (fig. 6) RF power level output, and set
(9) Adjust A14R21 (fig. 6) for a multimeter indication of 20 mV dc 2 mV.
(10) Press Range (F1) key, press DATA SCROLL arrow keys to select 200 W range.
Press DATA ENTRY ENTER key.
(11) Adjust A14R18 (fig. 6) for a multimeter indication of 10 mV dc 2 mV.
(12) Press Range (F1) and DATA SCROLL arrow keys to select 2 W range, and
press DATA ENTRY ENTER key.
(13) Remove the 50 Ω load from TI T/R.
(14) Connect measuring receiver sensor module to RF power amplifier OUTPUT
using 10 dB attenuator (fixed).
50