TB 9-6625-2296-24
(b) AUX (F6) ............................................. Auxiliary
(c) MEMORY RCL .................................. Recall
(d) FIELD SELECT ............................. Highlight 10. Factory Defaults
(e) DATA ENTRY ENTER ..................... Factory Defaults recall
NOTE
Assure highlighted area indicates YES before proceeding.
DATA ENTRY ENTER ...................... Factory Defaults
(f)
(2) Connect measuring receiver sensor module to TI T/R.
(3) Set measuring receiver to measure FM deviation, low pass filter 15 kHz, high
pass filter 50 Hz, and PEAK/2 detector selected.
(4) Press key sequence listed in (a) through (h) below:
(a) MODE RF GEN ................................................
RF Gen Display
(b) SETUP ...............................................................
Gen Menu
(c) DATA ENTRY 5 ................................................
RF Gen Setup
(d) DATA ENTRY 1 ................................................
RF Gen Freq
(e) DATA ENTRY 7 5 0 K......................................
750 kHz
(f) DATA ENTRY 2 ................................................
RF Gen Level
(g) DATA ENTRY 0 ENTER .................................
0 dBm
(h) Ret (F5) ..............................................................
Return
(5) Record measuring receiver indication as residual FM in table 18. Repeat this
step for remaining measuring receiver low pass filter settings listed in table 18 for TI RF
Measuring receiver
TI
Low pass filter
Residual FM
RF frequency
settings
indications
750 kHz
15
kHz
750 kHz
3
kHz
900 MHz
3
kHz
900 MHz
>20
kHz
(6) Repeat (4) (e) and (5) above for MODE RF GEN frequency of 900 MHz.
(7) Press (F4) SOURCE (Modulation Source) key. Press DATA SCROLL arrow
keys until SOURCE 1 MOD FM is displayed and press DATA ENTRY ENTER key.
57