TB 9-6625-2296-24
(a) MODE MTRS ..........................
Meter Menu
(b) AUX (F6) ..................................
Auxiliary Functions Menu
(c) DATA ENTRY 1 ......................
(d) FIELD SELECT arrow...........
Highlight 13. Gen Modulation
(e) DATA ENTRY ENTER ..........
Select 13. Gen Modulation
(f) DATA ENTRY 2 ......................
FM Deviation (20 kHz)
(g) DATA ENTRY 2, 0, 0, ENTER...... Correction Factor of 200
(h) DATA ENTRY ENTER................... Sets Correction Factor
(i) ESC (F6) ...................................
Escape
MODE RF GEN ......................
(j)
RF Gen Display
(k) DATA ENTRY SHIFT, Y .......
(Backup Cal Data ? y/n)
(5) Press key sequence listed in (a) through (q) below:
(a) MODE RF GEN ......................
RF Gen Display
(b) SETUP .....................................
Gen Menu
(c) DATA ENTRY 5 ......................
RF Gen Setup
(d) DATA ENTRY 1 ......................
RF Gen Freq
(e) DATA ENTRY 1 0 M ..............
10 MHz
(f) DATA ENTRY 2 ......................
RF Gen Level
(g) DATA ENTRY 0 ENTER .......
0 dB level
(h) ESC (F6) ...................................
Escape
(i) DATA ENTRY 1 ......................
Func Gen No. 1 Setup
DATA ENTRY 1 ......................
(j)
Func Gen 1
(k) DATA ENTRY 3 ......................
FM
(l) DATA ENTRY 2 ......................
Freq
(m) DATA ENTRY 2 2 6 4 . 5........
2264.5
(n) DATA ENTRY ENTER ..........
2264.5 Hz
(o) DATA ENTRY 4 ......................
FM Dev
(p) DATA ENTRY 1 2 . 5 K ..........
12.5 kHz
(q) Ret (F5) ....................................
Return
60