TB 9-6625-2296-24
a. Performance Check
(1) Press TI POWER key off and remove top and bottom covers.
(2) Press TI POWER key ON and allow 5 minutes for warm-up.
(3) Connect multimeter INPUT HI to orange wire on power termination and LO to
chassis ground.
(4) If multimeter does not indicate between +5.0 and +5.1 V dc, perform b (1) below.
(5) Connect multimeter INPUT HI to red wire and LO to ground lug (black wire on
(6) If multimeter does not indicate between +15.0 and +15.1 V dc perform b (2) below.
(7) Connect multimeter INPUT HI to yellow wire.
(8) Multimeter will indicate between -15.0 and -15.1 V dc.
(9) Connect multimeter INPUT HI to blue wire (fig. 6).
(10) Multimeter will indicate between +11.8 and +12.1 V dc.
(11) Connect multimeter INPUT HI to FL 16 (fig. 6) on 1st LO assembly.
(12) Multimeter will indicate between +33.0 and +35.0 V dc.
(13) Connect multimeter INPUT HI to A26TP2 (fig.2) and LO to A26TP3 (fig.2).
(14) If multimeter does not indicate +4.0960 0.0001 V dc, perform b (3) below.
(15) Disconnect multimeter from TI, press POWER key off, and reinstall top and
bottom covers. Press TI POWER key ON.
b. Adjustments
(1) Adjust A16R37 (fig. 2) for a multimeter indication of +5 V dc.
(2) Adjust A16R25 (fig. 2) for a multimeter indication of +15 V dc.
(3) Adjust A26R18 (fig. 2) for a multimeter indication of +4.0960 0.0001 V dc.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.