TB 9-6625-2296-35
Test instrument parameter
Performance specifications
RADIO FREQUENCY METER
Frequency Range
250 kHz to 999.9999 MHz
Level Range
-10 to 50 dBm (250 kHz to <200 MHz)
-10 to 47 dBm (200 to 999.9999 MHz)
Accuracy
0.5 PPM
Counter Range
0 to 150 kHz
Meter Range
0 to 100 kHz
Level Range
-10 to +50 dBm at T/R connector
Accuracy
0.5 ppm
AUDIO FREQUENCY METER
Range
10.0 Hz to 40.0 kHz
Accuracy
0.5 ppm
FM DEVIATION METER
0 Hz to 100 kHz (+ peak and - peak)
Range
Modulation Rate
100 Hz to <20 kHz (300 kHz IF)
100 Hz to <3 kHz (30 kHz IF)
Accuracy
4.0% 2 counts (300 kHz IF, <15 kHz rate) (FM zeroed)
7.0% 2 counts (300 kHz IF, ≥15 kHz rate) (FM zeroed)
7.0 %, 2 counts, (30 kHz IF) (FM zeroed)
AM MODULATION METER
Range
30% to 90%
Modulation Rate
100.0 Hz to 10.0 kHz
Accuracy
5% full scale, 1 count from 30% thru 90%
DISTORTION METER
1% to 20.0%
Range
0.5% distortion, 1 count, from 1% thru ≤10%
Accuracy
2.0% distortion, 1 count, from > 10.0% thru 20.0%
Frequency
700 Hz and 1 kHz
SINAD METER
770 Hz and 1.0 kHz
0.1 thru 30 V rms
Signal Level
3.0 thru 30.0 dB
Range
1.0 dB, 1 count
Accuracy
Frequency Range
10.0 Hz to 40.0 kHz
Accuracy
0.1%
0.7 mV rms to 2.5 V rms (150 Ω) up to 3.0 V rms (600 Ω)
Level range
(150 Ω load) 0.1 mV or <3% (≤10 kHz)
Accuracy
<5.0% (>10 to 25 kHz)
Spectral Purity THD (sine wave only)
<1.0%
<0.7% (0.7 to 1.1 kHz)
4