TB 9-6625-2296-35
NOTE
Assure highlighted area indicates YES before proceeding.
Factory Defaults restored
(11) DATA ENTRY ENTER ..............
Auxiliary Functions Menu
(12) AUX (F5)....................................
Self Test
(13) DATA ENTRY 4 .......................
Test Running
(14) DATA ENTRY ENTER............
NOTE
When self test is completed, ensure P is displayed next to Self
Test; otherwise refer to TM for maintenance action.
f. Press key sequence listed in (1) through (4) below:
Receiver screen
(1) MODE RCVR ..............................
Additional Selections
(2) More (F6) .....................................
121.1 MHz
(3) DATA ENTRY 1 2 1 . 1 M ..........
(4) Verify RF IN indicates ANT; if not, press Ant (F3) key.
g.
Connect signal generator RF OUTPUT to TI
ANTENNA.
h.
Set signal generator for a
121.1 MHz,
-30 dBm
output.
i. Press key sequence listed in (1) through (6) below:
Meter Menu
(1) MODE MTRS ..............................
Auxiliary Functions Menu
(2) AUX (F6) ......................................
(3) DATA ENTRY 1..........................
Signal Meter
(4) DATA ENTRY 7..........................
Enter to Cal
(5) DATA ENTRY ENTER ..............
Correction Factor
(6) DATA ENTRY 1 0 0 ENTER.....
j. Press Ret (F5) (Return), MODE RCVR, DATA ENTRY SHIFT, DATA ENTRY Y
(Backup Cal Data ? y/n) keys.
8. AF Generator Frequency
a. Performance Check
(1) Press key sequence listed in (a) through (f) below:
Meter Menu
(a) MODE MTRS ........................
Auxiliary Functions Menu
(b) AUX (F6) ................................