TB 9-6625-2296-35
(8) Press key sequence listed in (a) through (g) below:
Select GEN 1
(a) FIELD SELECT ⇑ ................
Set GEN 1 Off
(b)
DATA ENTRY ENTER ........
Select GEN 2
FIELD SELECT ⇓ ................
(c)
Set GEN 1 On
(d)
DATA ENTRY ENTER ........
Select GEN 1 AF
FIELD SELECT ⇓ ................
(e)
(f)
DATA ENTRY ENTER ........
20 Hz setting
(g)
DATA ENTRY 2 0 ENTER ..
(9) Repeat technique of (4) through (7) above for GEN 2.
(10) Press FIELD SELECT ⇑ key to select GEN 2 and press DATA ENTRY
ENTER key to set GEN 2 Off.
b. Adjustments
(1) Disconnect equipment from TI, press TI POWER key off, and remove function
generator assembly A24 (fig. 1). Install function generator assembly using extender board
(7010-7839-900). Press TI POWER key ON.
14