TB 9-6625-2296-35
(4) Press Range (F2) key, adjust DATA SCROLL knob to select TI range settings,
press DATA ENTRY ENTER key, and set calibrator for output settings listed in table 6.
TI multimeter indications will be within limits specified in table 6.
Test
Test instrument
instrument
Calibrator output settings
range settings
Amplitude
Min
Max
200 mV
180 mV
1 kHz
169.9
mV
190.1
mV
2.0 V
1.0 V
50 Hz
0.899 V
1.101
V
2.0 V
1.0 V
100 Hz
0.899 V
1.101
V
2.0 V
1.0 V
1 kHz
0.899 V
1.101
V
2.0 V
1.0 V
10 kHz
0.899 V
1.101
V
2.0 V
1.0 V
20 kHz
0.899 V
1.101
V
2.0 V
1.8 V
1 kHz
1.699 V
1.901
V
20 V
18 V
1 kHz
16.99
V
19.01
V
200 V
180 V
100 Hz
169.9
V
190.1
V
2000 V1
470 V
100 Hz
369
V
571
V
1After
verification press RESET key on calibrator.
(5) Press Func (F1) key, adjust DATA SCROLL knob to select DCV, and press
DATA ENTRY ENTER key.
(6) Press Range (F2) key, adjust DATA SCROLL knob to select TI range settings,
press DATA ENTRY ENTER key, and set calibrator for output settings listed in table 7.
TI multimeter indications will be within limits specified in table 7.
Test instrument
Test instrument
Calibrator
range settings
output settings
Min
Max
200
mV
190
mV
187.9
mV
192.1
mV
2.0
V
1.9
V
1.879
V
1.921
V
20
V
19
V
18.79
V
19.21
V
200
V
190
V
187.9
V
192.1
V
V1
2000
980
V
959
V
1001
V
1After
verification press RESET key on calibrator.
(7) Connect calibrator OUTPUT HI and LO to TI AMP and COM.
(8) Press Func (F1) key, adjust DATA SCROLL knob to select DCC, and press
DATA ENTRY ENTER key.
(9) Press Range (F2) key, adjust DATA SCROLL knob to select TI range settings,
press DATA ENTRY ENTER key, and set calibrator for output settings listed in table 8.
TI multimeter indications will be within limits specified in table 8.
21