TB 9-6625-2296-35
NOTE
Assure highlighted area indicates YES before proceeding.
Factory Defaults restored
(f) DATA ENTRY ENTER ..........
NOTE
If necessary, perform measuring receiver zero and calibrate.
(2) Connect measuring receiver sensor module to TI T/R IN/OUT.
(3) Press key sequence listed in (a) through (h) below:
RF Gen Display
(a) MODE RF GEN ......................
Gen Menu
(b) SETUP .....................................
RF Gen Setup
(c) DATA ENTRY 5 ......................
RF Gen Level
(d) DATA ENTRY 2 ......................
0 dBm
(e) DATA ENTRY 0 ENTER .......
RF Gen Freq
(f) DATA ENTRY 1 ......................
30 MHz
(g) DATA ENTRY 3 0 M ..............
Return
(h) Ret (F5) ....................................
(4) Set measuring receiver to measure RF power at 30 MHz.
(5) Set measuring receiver to ratio mode.
(6) Press TI DATA SCROLL arrow keys to manually sweep TI frequency from
500 kHz to 399 MHz entering TI display frequency into measuring receiver. If measuring
receiver indication does not remain within 2.0 dB, perform b below.
(7) Press TI DATA SCROLL arrow keys to manually sweep TI frequency from
400 MHz to 999 MHz entering TI display frequency into measuring receiver. If measuring
receiver indication does not remain within 3.5 dB, perform b below.
(8) Set TI frequency to 30 MHz. Set measuring receiver to acquire frequency then
to measure tuned rf power and perform measuring receiver calibrate.
NOTE
(9) Press TI Level (F3) (dBm display) key and adjust output level for the following
dBm indications : 0, -1, -2, -4, -8, -16, -32 and -64. Measuring receiver indications will be
within 2.0 dBm of TI indications.
(10) Adjust output level from -90 dBm to -110 dBm in 10 dB steps. Measuring
receiver indications will be within 2.5 dBm of TI indications.
23