TB 9-6625-2296-35
(29) Set frequency to 935.02 MHz on spectrum analyzer and wait for several sweeps.
Spectrum analyzer will indicate <-85 dB (1 Hz).
b. Adjustments. No further adjustments can be made.
a. Performance Check
(1) Press key sequence listed in (a) through (f) below:
Meters Menu
(a) MODE MTRS ..........................
Auxiliary Functions Menu
(b) AUX (F6) ..................................
Recall Parameters Menu
(c) MEMORY RCL .......................
(d) FIELD SELECT ⇑ ..................
Highlight 10. Factory Defaults
Factory Defaults recall
(e) DATA ENTRY ENTER ..........
NOTE
Assure highlighted area indicates YES before proceeding.
Factory Defaults recall
(f) DATA ENTRY ENTER ..........
(2) Connect measuring receiver sensor module to TI T/R IN/OUT.
(3) Set measuring receiver as listed in (a) through (e) below:
(a) INSTR PRESET.
(b) FM deviation.
(c) 50 Hz HIGH PASS filter.
(d) 15 kHz LOW PASS filter.
(e) RMS detector.
(4) Press key sequence listed in (a) through (h) below:
RF Gen Display
(a) MODE RF GEN ......................
Gen Menu
(b) SETUP .....................................
RF Gen Setup
(c) DATA ENTRY 5 ......................
RF Gen Level
(d) DATA ENTRY 2 ......................
0 dBm
(e) DATA ENTRY 0 ENTER .......
RF Gen Freq
(f) DATA ENTRY 1 ......................
1 MHz
(g) DATA ENTRY 1 M .................
Return
(h) Ret (F5) ....................................
(5) Measuring receiver FM indication will be <45 Hz.
(6) Enter 9 9 . 9 9 9 9 M using TI DATA ENTRY keys.
(7) Measuring receiver FM indication will be <45 Hz.
(8) Enter 5 0 0 M using TI DATA ENTRY keys.