TB 9-6625-2296-35
Test instrument
frequency settings
Frequency counter indications
(MHz)
Min
Max
0.500
499.99975 kHz
500.00025
kHz
1.0
999.9995
kHz
1.0000005
MHz
10.0
9.999995 MHz
10.000005
MHz
100.0
99.999950 MHz
100.000050
MHz
1
500.02
499.999750 MHz
500.000250
MHz
999.0
998.999500 MHz
999.000500
MHz
1Set
frequency counter A input to 50 Ω.
2Move connection from frequency counter A input to C input.
b. Adjustments
(1) Press key sequence listed in (a) through (f) below:
Meters Menu
(a) MODE MTRS ..........................
Auxiliary Function Menu
(b) AUX (F6) ..................................
Recall
(c) MEMORY RCL .......................
(d) FIELD SELECT ⇑ ..................
Highlight 10. Factory Defaults
Factory Defaults recall
(e) DATA ENTRY ENTER ..........
NOTE
Assure highlighted area indicates YES before proceeding.
Factory Defaults restored
(f) DATA ENTRY ENTER ..........
(2) Connect frequency counter C input to TI T/R IN/OUT.
(3) Press key sequence listed in (a) through (h) below:
RF Gen Display
(a) MODE RF GEN ......................
Gen Menu
(b) SETUP .....................................
RF Gen Setup
(c) DATA ENTRY 5 ......................
RF Gen Level
(d) DATA ENTRY 2 ......................
0 dBm
(e) DATA ENTRY 0 ENTER .......
RF Gen Freq
(f) DATA ENTRY 1 ......................
900 MHz
(g) DATA ENTRY 9 0 0 M ...........
Return
(h) Ret (F5) ....................................