TB 9-6625-2296-35
Auxiliary Function Menu
(b) AUX (F6) ..................................
Recall
(c) MEMORY RCL .......................
(d) FIELD SELECT ⇑ ..................
Highlight 10. Factory Defaults
Factory Defaults recall
(e) DATA ENTRY ENTER ..........
NOTE
Assure highlighted area indicates YES before proceeding.
Factory Defaults restored
(f) DATA ENTRY ENTER ..........
Analyzer Display
(g) SCOPE/ANLZ twice................
(12) Press key sequence listed in (a) through (n) below:
Scope/Analyzer Menu
(a) SETUP .....................................
Analyzer On
(b) DATA ENTRY 2 ......................
Setup Analyzer
(c) DATA ENTRY 4 ......................
(d) DATA ENTRY 2 ......................
1 MHz
(e) DATA ENTRY 1 M .................
Input Atten
(f) DATA ENTRY 4 ......................
40 dB
(g) DATA ENTRY 3 ......................
Scan Width
(h) DATA ENTRY 6 ......................
DATA SCROLL ⇓.............................
Highlight 5 kHz
(i)
Select 5 kHz Scan Width
(j)
DATA ENTRY ENTER ..........
(k) FIELD SELECT ⇓............................
Highlight 10. Velocity Factor
Select Velocity Factor
(l) DATA ENTRY ENTER ..........
100%
(m) DATA ENTRY 1 0 0 ENTER ........
Return
(n) Ret (F5) ....................................
(13) Disconnect signal generator from TI ANTENNA. Press Norm (F1) key and
allow time for normalize cycle to complete.
(14) Connect signal generator RF OUPUT to variable attenuator input. Connect
variable attenuator output to TI ANTENNA.
amplitude for a 0 dB TI indication.
(16) Set variable attenuator from 0 to 40 dB in 10 dB steps. TI indication will be
within 4 dB of variable attenuator test report values.
(17) Press Atten (F4) (RF ATTEN) key and press DATA SCROLL ⇑ key until RF
ATTEN indicates 0 dB and repeat (13) and (14) above.
35