TB 9-6625-2296-35
Input Atten
(e) DATA ENTRY 4 ......................
40 dB
(f) DATA ENTRY 3 ......................
Return
(g) Ret (F5) ....................................
amplitude for a 0 dB TI indication.
(Normalize) key. Reconnect after allowing time for normalize cycle to complete. If
necessary, adjust signal generator amplitude for a 0 dB TI indication.
(30) Repeat technique of steps (16) through (18) above. TI indications will be within
5 dB of variable attenuator test report values.
(31) Disconnect signal generator and variable attenuator from TI.
(32) Press key sequence listed in (a) through (g) below:
Meters Menu
(a) MODE MTRS ..........................
Auxiliary Function Menu
(b) AUX (F6) ..................................
Recall
(c) MEMORY RCL .......................
(d) FIELD SELECT ⇑ ..................
Highlight 10. Factory Defaults
Factory Defaults recall
(e) DATA ENTRY ENTER ..........
NOTE
Assure highlighted area indicates YES before proceeding.
Factory Defaults restored
(f) DATA ENTRY ENTER ..........
Analyzer Display
(g) SCOPE/ANLZ twice................
(33) Press key sequence listed in (a) through (j) below:
Scope/Analyzer Menu
(a) SETUP .....................................
Analyzer On
(b) DATA ENTRY 2 ......................
Setup Analyzer
(c) DATA ENTRY 4 ......................
(d) DATA ENTRY 2 ......................
121.5 MHz
(e) DATA ENTRY 1 2 1 . 5 M ......
Range 2 dB
(f) DATA ENTRY 5 ......................
Scan Width
(g) DATA ENTRY 6 ......................
(h) DATA SCROLL ⇓ ...................
Highlight 100 kHz
Select 100 kHz Scan Width
(i) DATA ENTRY ENTER ..........
Return
(j)
Ret (F5) ....................................
(34) Connect signal generator RF OUPUT to ANTENNA and set signal generator
for a 121.5 MHz, -40 dBm output.
(35) Press (F4) (Ref lvl) and DATA SCROLL arrow keys to place signal peak to
approximately center of display.
37