TB 9-6625-2296-35
a. Performance Check
(1) Press key sequence listed in (a) through (f) below:
Meters Menu
(a) MODE MTRS ..........................
Auxiliary Function Menu
(b) AUX (F6) ..................................
Recall
(c) MEMORY RCL .......................
(d) FIELD SELECT ⇑...................
Highlight 10. Factory Defaults
Factory Defaults recall
(e) DATA ENTRY ENTER ..........
NOTE
Assure highlighted area indicates YES before proceeding.
Factory Defaults restored
(f) DATA ENTRY ENTER ..........
(2) Press MODE MTRS (Meters Menu) key and DATA ENTRY 3 key (Pwr Meter
(Pulse/CW).
Figure 6. Power meter setup (20mW 500mW range).
(4) Press Range (F1) and DATA SCROLL arrow keys to select 20 mW range and
press Zero (F4) key. Press DATA ENTRY ENTER key when Zero submenu appears.