TB 9-6625-2296-35
power amplifier output until measuring receiver indicates exactly 33 dB minus 10 dB
attenuator (fixed) test report value.
(29) Disconnect RF power amplifier and 10 dB attenuator (fixed) from measuring
receiver. Connect RF power amplifier OUTPUT to TI T/R using 10 dB attenuator (fixed).
(30) Press Range (F1) key repeatedly to select 2 W range, and press DATA ENTRY
ENTER key.
(31) Press key sequence listed in (a) through (e) below:
Auxiliary Function Menu
(a) AUX (F6) ..................................
(b) DATA ENTRY 1 ......................
Power Meter
(c) DATA ENTRY 1 ......................
Power entry mode
(d) DATA ENTRY ENTER ..........
2W
(e) DATA ENTRY 2 ENTER .......
(32) Press Range (F1) (power meter range mode) repeatedly to select 200 W range
and press DATA ENTRY ENTER (200 W range) key.
power amplifier output to 50W.
(35) Press key sequence listed in (a) through (f) below:
Auxiliary Functions Menu
(a) AUX (F6) ..................................
(b) DATA ENTRY 1 ......................
Power Meter
(c) DATA ENTRY 1 ......................
ENTER
(d) DATA ENTRY ENTER ..........
50 W
(e) DATA ENTRY 5 0 ENTER ....
Return
(f) Ret (F5) ....................................
(36) Press MODE RF GEN (MODE RF GEN Display); when prompted for backup
data, press SHIFT and Y keys (Backup Cal Data ? y/n).
(37) Disconnect all test equipment.
20. Generator Amplitude Modulation
a. Performance Check
(1) Press key sequence listed in (a) through (f) below:
Meters Menu
(a) MODE MTRS ..........................
Auxiliary
(b) AUX (F6) ..................................
Recall
(c) MEMORY RCL .......................
(d) FIELD SELECT ⇑...................
Highlight 10. Factory Defaults