TB 9-6625-2296-35
Return
(f) Ret (F6) ....................................
FM Zero
(g) FM Z (F3) .................................
(8) Press MODE MTRS (Meters Menu), AUX (F6) (Auxiliary Functions Menu), and
DATA ENTRY ENTER keys, then press FIELD SELECT arrow keys to highlight 2. Dev
Meter (Peak). Press DATA ENTRY ENTER key.
(9) Press DATA ENTRY ENTER and DATA ENTRY keys to enter value displayed
on measuring receiver in 10 kHz(+) data field and press DATA ENTRY ENTER key.
(10) Press DATA ENTRY - key, followed by DATA ENTRY numeric keys to enter
value displayed on measuring receiver in 10 kHz(-) data field and press DATA ENTRY
ENTER key.
(11) Adjust signal generator output for 121.1 MHz at 0 dBm with 17 kHz deviation at
3 kHz rate. Set FM off.
(12) Press Ret (F5) (Return), MODE RCVR (Receiver Display), SHIFT, and Y keys
(Backup Cal Data ? y/n).
(13) Press key sequence listed in (a) through (g) below:
(a) DATA SCROLL arrow keys ...
Highlight DEV
(b) DATA ENTRY ENTER ..........
Select DEV
Deviation Range
(c) Range (F1) ...............................
(d) DATA SCROLL arrow keys ...
Highlight 20 kHz range
Select 20 kHz Deviation
(e) DATA ENTRY ENTER ..........
Return
(f) Ret (F6) ....................................
FM Zero
(g) FM Z (F3) .................................
(15) Press MODE MTRS (Meters Menu), AUX (F6) (Auxiliary Functions Menu), and
DATA ENTRY ENTER keys; press FIELD SELECT arrow keys to highlight 2. Dev
Meter (Peak) and press DATA ENTRY ENTER key.
(16) Press DATA ENTRY ENTER and DATA ENTRY keys to enter value displayed
on measuring receiver in 20 kHz(+) data field, and press DATA ENTRY ENTER key.
(17) Press DATA ENTRY - key, followed by DATA ENTRY numeric keys to enter
value displayed on measuring receiver in the 20 kHz(-) data field, and press DATA ENTRY
ENTER key.
(18) Press Ret (F5) (Return), MODE RCVR (Receiver Display), SHIFT, and Y keys
(Backup Cal Data ? y/n).
60