TB 9-6625-2296-35
(19) Adjust signal generator output for 121.1 MHz at 0 dBm with 45 kHz deviation at
3 kHz rate and set FM off.
(20) Press key sequence listed in (a) through (g) below:
(a) DATA SCROLL arrow keys ...
Highlight DEV
(b) DATA ENTRY ENTER ..........
Select DEV
Deviation Range
(c) Range (F1) ...............................
(d) DATA SCROLL arrow keys ...
Hightlight 50 kHz range
Select 50 kHz Deviation
(e) DATA ENTRY ENTER ..........
Return
(f) Ret (F6) ....................................
FM Zero
(g) FM Z (F3) .................................
(21) Set signal generator FM on.
(22) Press MODE MTRS (Meters Menu), AUX (F6) (Auxiliary Functions Menu), and
DATA ENTRY ENTER keys; press FIELD SELECT arrow keys to highlight 2. Dev
Meter (Peak) and press DATA ENTRY ENTER key.
(23) Press DATA ENTRY ENTER key and press DATA ENTRY keys to enter value
displayed on measuring receiver in 50 kHz(+) data field. Press DATA ENTRY ENTER key.
(24) Press DATA ENTRY - key, followed by DATA ENTRY numeric keys to enter
value displayed on measuring receiver in 50 kHz(-) data field, and press DATA ENTRY
ENTER key.
(25) Adjust signal generator output for 121.1 MHz at 0 dBm with 75 kHz deviation at
3 kHz rate and set FM off.
(26) Press Ret (F5) (Return), MODE RCVR (Receiver Display), SHIFT, and Y keys
(Backup Cal Data ? y/n).
(27) Press key sequence listed in (a) through (g) below:
(a) DATA SCROLL arrow keys ...
Highlight DEV
(b) DATA ENTRY ENTER ..........
Select DEV
Deviation Range
(c) Range (F1) ...............................
(d) DATA SCROLL arrow keys ...
Hightlight 100 kHz range
100 kHz deviation
(e) DATA ENTRY ENTER ..........
Return
(f) Ret (F6) ....................................
FM Zero
(g) FM Z (F3) .................................
(28) Set signal generator FM on.
(29) Press MODE MTRS (Meters Menu), AUX (F6) (Auxiliary Functions Menu), and
DATA ENTRY ENTER keys; press FIELD SELECT arrow keys to highlight 2. Dev
Meter (Peak) and press DATA ENTRY ENTER key.
61