TB 9-6625-2296-35
Meter Menu
(m) MODE MTRS ..........................
Freq Error Meter/Freq Meter
(n) DATA ENTRY 2 ......................
(17) Set signal generator for a 900.15 MHz, 0 dBm output.
900.15000 MHz.
(19) Frequency counter will indicate 900.15000 MHz 450.08 Hz. TI FREQ ERR
display will indicate 150 kHz.
899.85000 MHz.
(21) Frequency counter will indicate 899.85000 MHz 449.93 Hz. TI FREQ ERR
display will indicate 150.0 kHz.
b. Adjustments. No further adjustments can be made.
a. Performance Check
(1) Press key sequence listed in (a) through (f) below:
Meters Menu
(a)
MODE MTRS ..........................
Auxiliary
(b)
AUX (F6) ..................................
Recall
(c)
MEMORY RCL .......................
FIELD SELECT ⇑ ..................
Highlight 10. Factory Defaults
(d)
Factory Defaults recall
(e) DATA ENTRY ENTER............
NOTE
Assure highlighted area indicates YES before proceeding.
Factory Defaults restored
(f) DATA ENTRY ENTER ..........
(2) Press MODE RCVR (Receiver Display), Freq (F1) (Frequency), DATA ENTRY
1 and M (1 MHz) keys.
(3) Connect frequency counter 10 MHz OUT to function/arbitrary generator 10
MHz In.
(4) Connect function/arbitrary generator Output to TI EXT MOD IN and frequency
counter A input.
(5) Press key sequence listed in (a) through (e) below:
(a) FIELD SELECT arrow key ....
Highlight AF
AUDIO FREQUENCY DISPLAY
(b) DATA ENTRY ENTER ..........
Input
(c) F4 ..............................................