TB 9-6625-2308-35
(11) Using the formula below, calculate KB.
Pt
x 100
KB =
Pi
(12) If TI displayed power is not within 1.61% of calculated Pt , perform b below.
(13) Repeat technique of (2) through (11) above for frequencies and tolerances listed
in table 4. Make appropriate changes to sensor CALFAC as required using technique in b below.
(GHz)
Tolerance
1.95%
2.0
2.44%
3.0
2.44%
4.0
2.67%
5.0
2.67%
6.0
2.86%
7.0
2.86%
8.0
3.59%
9.0
3.59%
10.0
3.59%
11.0
3.59%
12.0
4.09%
13.0
4.09%
14.0
4.09%
15.0
4.09%
16.0
4.09%
17.0
4.09%
18.0
b. Adjustments
(1) Press TI MENU, SERVICE, SENSOR ROM softkeys.
(2) Scroll cursor to appropriate CALFAC and change to KB calculated in (11) above.
(3) Press ENTER.
a. Deenergize and disconnect all equipment and reinstall protective covers on TI if necessary.
b. Annotate and affix DA label/form in accordance with TB 750-25.
PIN: 075096-001