TB 9-6625-2308-40
Table 2. Minimum Specifications of Equipment Required Continued.
Manufacturer and model
Common name
Minimum use specifications
(part number)
PNA NETWORK
Frequency range: 10 MHz to 50 GHz
E8364C with E-Cal modules
ANALYZER
(dB)
N4690-60004, N4691-60004
45 MHz to 2 GHz
0.001
and N4693-60001
2 to 20 GHz
0.008
20 to 40 GHz
0.020
40 to 50 GHz
0.020
Aeroflex, Model 2023B
Power output: 0 to +8 dBm
(2023B OPT2/4/11/122)
Accuracy: 0.8 dBm
STD RF XFR NO. 1
Frequency range: 10 MHz to 18 GHz
Tegam, Model F1109
Accuracy:
0.01 to 10 GHz, 1.00%
10 to 18 GHz, 1.10%
STD RF XFR NO. 2
Tegam, Model F1116
Accuracy:
0.1 to 10 MHz, 0.80%
10 to 100 MHz, 0.90%
STD RF XFR NO. 3
Frequency range: 10 MHz to 26.5 GHz
Tegam, Model F1135
Accuracy:
0.01 to 10 GHz, 1.00%
10 to 18 GHz, 1.10%
18 to 26.5 GHz, 2.20%
SYNTHESIZED
Frequency range: 2 MHz to 26.5 GHz
Anritsu, Model 68369NV
(68369NV)
Accuracy: 2 dB
Two needed.
1
POWER METER, GIGA-TRONICS MODEL 8542
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.