TB 9-6625-2309-35
(c) From selected TI optical module data sheet, type in data from Power Low
Ref (dBm) column and press ENTER.
(d) Repeat technique of (b) and (c) above as necessary to enter remaining data
from data sheet.
(e) When all data has been entered press Esc.
(41) Perform steps (a) through (i) below:
(a) Press 2 on key board to access Enter/Edit Attenuation Measurement Data.
(b) Press I on keyboard to insert data.
(c) At prompt, Enter power high (dBm):, type in data from Power High
column on selected TI optical module data sheet (appendix A).
(d) At prompt, Enter power low (dBm):, type in data from Power Low data
column on selected TI optical module data sheet (appendix A).
(e) Repeat technique of steps (b) through (d) above as necessary for remaining
data on TI optical module data sheet.
(f) Verify data has been entered correctly.
(g) Press S on keyboard at prompt and enter B = reference cursor position in
km.
(h) Press Q on keyboard to quit data edit page.
(i) OFTS Attenuation Scale Calibration menu will appear on screen.
(42) Perform steps (a) through (c) below:
(a) Press 3 on keyboard to select Generate Report.
(b) Generate Attenuation Report menu will appear on screen.
(c) Press 2 on keyboard to select Graphic Report (Screen) then press 1 on
keyboard to select Graphic Report (Printer) to print a hard copy to be sent to customer.
Refer to note immediately following figure 14 for pass/fail condition.
24