TB 9-6625-2309-35
(11) Adjust CURSORS A control to highlight filename OTDRCODE.EXE 03/13/95.
(12) Press SELECT FILE softkey.
(13) Adjust CURSORS A control to highlight filename OTSFTXTC.DAT 03/13/95.
(14) Press SELECT FILE softkey.
(15) Press COPY SELECTIONS softkey.
(16) Press C: key on keyboard.
NOTE
CURSORS A & B knobs can be used to enter drive letter.
Keyboard is needed for a warm boot of TI to maintain correct
(17) Ensure C: appears just below Copy Selected File(S) to: on TI display.
(18) Press START COPY softkey.
(19) Wait for message Copying in Progress PLEASE WAIT to disappear from TI
display.
(20) Remove floppy disk from external drive.
(21) On keyboard simultaneously press keys Ctrl, Alt, Del to warm boot TI.
(22) After warm boot of TI, Software Level 1.08a should appear when Press Any
Main Panel Key To Begin screen is displayed.
NOTE
Steps in c below are used to move optical trace to left or right
causing a change in distance scale reading.
NOTE
If steps in c below are used to change ZERO-KM SETTING to a
value other than 0, the ZERO-KM SETTING must be returned
to 0 for steps in Section III, paragraph 7a and b to be valid.
Record value of ZERO-KM SETTING and restore this value
c. To change ZERO-KM SETTING perform steps (1) through (10) below:
(1) If Press Any Main Panel Key To Begin screen is displayed press any key,
then press HELP pushbutton or press HELP pushbutton if distance scale is displayed on
crt.
34