TB 9-6625-2311-24
Table 2. Minimum Specifications of Equipment Required - Continued
Manufacturer and model
Common name
Minimum use specifications
(part number)
MEASURING RECEIVER
Frequency range: 100 kHz to 1.2 GHz
Measuring receiver system
mV range: 100 mV rms
N5530S consisting of:
Accuracy: +.258%
Spectrum Analyzer, Agilent
Model E4440A (E4440A),
Power meter, Agilent Model
E4419B (E4419B), and Sensor
module, Agilent Model 504
(504)
Hewlett-Packard, Model
Range: -15.2 to 15.2 V dc
Accuracy: +.038%
3458A (3458A)
POWER SPLITTER
Weinschel, Model 1870A (7916839)
(Aeroflex, Model 2023B
Output amplitude: 100 mV
(2023B) or SG-1207/U (SG-
1207/U)
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the
calibration process. Personnel should become familiar with the entire bulletin
before beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by
c. Unless otherwise specified, verify the result of each test and, whenever the
test requirement is not met, take corrective action before continuing with the
calibration. Adjustments required to calibrate the TI are included in this procedure.
Additional maintenance information is contained in the manufacturer's manual for
this TI.
d. When indications specified in paragraph 8 are not within tolerance, perform
the power supply check prior to making adjustments. After adjustments are made,
repeat paragraph 8. Do not perform power supply check if all other parameters are
within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.