TB 9-6625-2311-24
(22) Set up measuring receiver to measure RF power in mV.
(24) Adjust signal generator amplitude controls for a 100.0 mV indication on
measuring receiver. Record TI indication under TI instrument indications No. 1
(25) Repeat technique of (23) and (24) above for remaining signal generator
frequency settings listed in table 3 below.
(26) Set signal generator output to off.
(27) Reverse the output port connections of the power splitter.
(28) Set signal generator output to on.
(29) Repeat technique of (23) and (24) above and record TI indication under TI
indications No. 2 column in table 3 below.
NOTE
(30) Add TI indications No. 1 value to TI indications No. 2 value and divide
the sum by two. Record results in TI indications average column. Results recorded
in TI indications average column will be within the limits specified in the minimum-
(31) Repeat (30) above for each row listed in table 3.
Test instrument indications
Signal generator
settings
No. 1
No. 2
Average
Min
Max
(MHz)
(mV)
(mV)
(mV)
(mV)
(mV)
10.00
97.9
102.1
30.00
97.9
102.1
60.00
97.9
102.1
80.00
97.9
102.1
90.00
97.9
102.1
110.00
95.9
104.1
200.00
95.9
104.1
400.00
95.9
104.1
600.00
95.9
104.1
800.00
95.9
104.1
900.00
95.9
104.1
1100.00
91.9
108.1
1200.00
91.9
108.1
7