TB 9-6625-2315-24
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Sets AN/GSM-286, AN/GSM-287 or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Dc voltage:
Fluke, Model 5720A (5720A) (p/o MIS-
Range: 100 mV to 1000 V
35947); w amplifier, Fluke 5725A/AR
(5725A/AR)
Accuracy:
0.001%
Range
Accuracy
0.001
0
0.002 %
100 to 1 M
0.010 %
10 M
0.200 %
100 M
Dc current:
Range
Accuracy
0.5 A
0 mA
10 mA
0.017 %
100 mA
0.013 %
1A
0.027 %
2A
0.037 %
Ac voltage:
Range: 100 mV to 750 V
Frequency
Accuracy
0.022 %
10 Hz to 20 kHz
0.042 %
20 to 50 kHz
0.17 %
100 kHz
1.12 %
300 kHz
Ac Current:
.035 %
1A
1 kHz
.052 %
2A
1 kHz
FUNCTION/ARBITRARY
Function: Sinewave
Agilent, Model 33250A (33250A)
Amplitude: 1 Vrms
Accuracy: 0.0025 %