TB 9-6625-2315-24
key until PARAMETERS is displayed.
(9) Press RANGE/DIGITS
or
and RANGE/DIGITS
or
keys until +100.000 mA
(10) Press MENU
DC is displayed.
(11) Press RANGE/DIGITS Auto/Man key. Adjustment takes approximately 20
seconds to complete then TI will return to measurement mode.
or
(12) Press RANGE/DIGITS
key to select 1 A range.
(13) Set calibrator for a 1 A output and repeat (5) through (9) above.
(14) Press MENU or
and RANGE/DIGITS
or
keys until +1.00000 A DC is
displayed. Repeat (11) above.
or
(15) Press RANGE/DIGITS
key to select 3 A range.
(16) Set calibrator for a 2 A output and repeat (5) through (9) above.
(17) Press MENU or
and RANGE/DIGITS
or
keys until +2.00000 A DC is
displayed. Repeat (11) above.
(18) Press calibrator RESET key.
e. Perform ac volt gain adjustments as listed in (1) through (13) below:
(1) Connect calibrator OUTPUT HI and LO to TI Input HI and LO.
(2) Press keys as listed in (a) through (e) below:
FUNCTION AC V.
(a)
Shift then MENU .
(b)
RANGE/DIGITS until AC FILTER is displayed.
(c)
MENU or until SLOW 3 HZ is displayed.
(d)
RANGE/DIGITS Auto/Man.
(e)
or
(3) Press RANGE/DIGITS
key to select 100 mV range.
(4) Set calibrator for a 100 mV, 1 kHz output.
key.
(5) Press Shift key then MENU
or
key until F: CAL MENU is displayed.
(6) Press MENU
key until 1: UNSECURED is displayed.
(7) Press RANGE/DIGITS
key until 2: CALIBRATE is displayed.
(8) Press MENU
key until PARAMETERS is displayed.
(9) Press RANGE/DIGITS
or
and RANGE/DIGITS
or
keys until 100.000 mV
(10) Press MENU
AC is displayed.
(11) Press RANGE/DIGITS Auto/Man key. Adjustment takes approximately 20
seconds to complete then TI will return to measurement mode.
(12) Repeat technique of (3) through (11) above for remaining ac volt adjustments
13