TB 9-6625-2315-35
Shift then MENU <.
(b)
RANGE/DIGITS ∨ until AC FILTER is displayed.
(c)
MENU < or > until SLOW 3 HZ is displayed.
(d)
(e)
RANGE/DIGITS Auto/Man.
(3) Press RANGE/DIGITS ∧ or ∨ key to select 100 mV range.
(4) Set calibrator for a 100 mV, 1 kHz output.
(5) Press Shift key then MENU < key.
(6) Press MENU < or > key until F: CAL MENU is displayed.
(7) Press RANGE/DIGITS ∨ key until 1: UNSECURED is displayed.
(8) Press MENU > key until 2: CALIBRATE is displayed.
(9) Press RANGE/DIGITS ∨ key until PARAMETERS is displayed.
(10) Press MENU < or > and RANGE/DIGITS ∧ or ∨ keys until 100.000 mV AC is
displayed.
(11) Press RANGE/DIGITS Auto/Man key. Adjustment takes approximately 20
seconds to complete then TI will return to measurement mode.
(12) Repeat technique of (3) through (11) above for remaining ac volt adjustments
Test instrument
Calibrator
Value displayed in
ranges
output
(10) above
Voltage
100 mV
100 mV
50 kHz
100.000
mV AC
1V
1
V
1 kHz
1.00000 V AC
10 V
10
V
1 kHz
10.0000
V AC
10 V
10
V
50 kHz
10.0000
V AC
10 V
10
V
10 Hz
10.0000
V AC
100 V
100
V
1 kHz
100.000
V AC
100 V
100
V
50 kHz
100.000
V AC
750 V
750
V
1 kHz
750.00
V AC
750 V
200
V
50 kHz
200.00
V AC
(13) Press calibrator RESET key and disconnect from TI.
f. Perform frequency adjustments as
listed in
(1) through (12) below:
using 50 Ω feedthrough termination.
(2) Press FUNCTION Freq key.
(3) Press RANGE/DIGITS ∧ or ∨ to select 1 V range.
output.
(5) Press Shift key then MENU < key.
14