TB 9-6625-2316-35
9. Insulation Resistance (100 kΩ)
a. Performance Check
(1) Set FUNCTION switch to IR (0.1M).
(2) Set AUTO/SINGLE switch to SINGLE.
(3) Press the RESET pushbutton.
(4) Set resistance standard No. 2 for 102 kΩ.
(5) Press the ADVANCE pushbutton once to test circuit A01. TI TEST light will
be on and the FAULT light will remain off; if not, perform b below.
(6) Press the RESET pushbutton and set resistance standard No. 2 for 98 kΩ.
(7) Press the ADVANCE pushbutton once to test circuit A01. TI FAULT light will
be on; if not, perform b below.
(8) Press the RESET pushbutton and disconnect resistance standard No.
b. Adjustments. Perform 8b above, if not previously performed. Otherwise, no other
adjustments can be made.
10. Insulation Resistance (10 MΩ)
a. Performance Check
(1) Set FUNCTION switch to IR (10M).
(2) Set AUTO/SINGLE switch to SINGLE.
(3) Press the RESET pushbutton.
(4) Connect resistance standard No. 3 in series with resistance standard No. 2 and
set for a total resistance value of 10.3 MΩ.
(5) Press the ADVANCE pushbutton once to test circuit A01. TI TEST light will
be on and the FAULT light will remain off; if not, perform b below.
(6) Press the RESET pushbutton and set resistance standard No. 2 and No. 3 for a
total resistance of 9.7 MΩ.
(7) Press the ADVANCE pushbutton once to test circuit A01. TI FAULT light will
be on; if not, perform b below.
7