TB 9-6625-2317-40
(d) LEVEL UNITS ABS.
(e) HIGH PASS FILTER 50 HZ.
(f) LOW PASS FILTER 15 kHz.
(4) Adjust signal generator frequency to 150 kHz at 0 dBm as indicated on
frequency counter.
(5) Press TI pushbuttons CARR FREQ, . (decimal), 1, 5, 0, & MHz.
(6) TI front panel FREQUENCY display will indicate between 149.97 and
150.03 kHz RF.
(7) Repeat techniques (4) through (6) above, using settings in table 3. TI frequency
will indicate within limits specified.
(8) On signal generator press RF OFF/ON pushbutton to OFF.
NOTE
TI may automatically select carrier frequency.
Signal Generator
Test instrument indications
(0 dBm)
Min
Max
500
kHz
499.97 kHz
500.03 kHz
1
MHz
999.97 kHz
1.00003 MHz
10
MHz
9.99996 MHz
10.00004 MHz
50
MHz
49.99992 MHz
50.00008 MHz
100
MHz
99.99987 MHz
100.00013 MHz
200
MHz
199.99977 MHz
200.00023 MHz
MHz1
500
499.99947
MHz
500.00053
MHz
1
GHz
999.9987
MHz
1000.0013
MHz
1.5
GHz
1499.9982
MHz
1500.0018
MHz
1On
frequency counter, move coax cable to C input. Set counter FUNCTION for Freq C with 50 Ω impedance.
6