TB 9-6625-2317-50
Test instrument indications
(0 dBm)
Min
Max
500
kHz
499.97
kHz
500.03
kHz
1
MHz
999.97
kHz
1.00003
MHz
10
MHz
9.99996
MHz
10.00004
MHz
50
MHz
49.99992
MHz
50.00008
MHz
100
MHz
99.99987
MHz
100.00013
MHz
200
MHz
199.99977
MHz
200.00023
MHz
500
MHz1
499.99947
MHz
500.00053
MHz
1
GHz
999.9987
MHz
1000.0013
MHz
1.5
GHz
1499.9982
MHz
1500.0018
MHz
1 On
frequency counter, move coax cable to C input. Set counter FUNCTION for Freq C with 50 Ω impedance.
b. Adjustments. No adjustments can be made.
a. Performance Check
(2) Set calibrator WIDEBAND output for 30 MHz frequency and amplitude of
+12 dBm.
(3) Press TI pushbuttons CARR FREQ, 3, 0, & MHz.
(4) Press TI LEVEL DISPLAY RF POWER pushbutton (on). TI LEVEL/DATA
ENTRY display will indicate between 9 and 15 dBm.
(5) Press calibrator OPR/STBY pushbutton to STBY.
b. Adjustments. No adjustments can be made.
11. Carrier
Level Sensitivity
a. Performance Check