TB 9-6625-2320-35
(30) Verify trigger READY indicator on the front-panel flashes about once every
second as the waveform is updated on-screen.
(31) Disconnect probe tip from PROBE COMPENSATION SIGNAL; disconnect
probe ground from PROBE COMPENSATION GND.
a. Performance Check
(1) Press front-panel SETUP key.
(2) Press main-menu Recall Factory Setup key.
(3) Press side-menu OK Confirm Factory Init key.
(4) Press front-panel CLEAR MENU key.
(5) Press front-panel SHIFT key.
(6) Press front-panel ACQUIRE MENU key.
(7) Press main-menu MODE key.
(8) Press side-menu Hi Res key.
(9) Press front-panel CURSOR key.
(10) Press main-menu Function key.
(11) Press side-menu H Bars key.
(12) Press front-panel CLEAR MENU key.
(13) Press front-panel CH 1 key.
(14) Press front-panel VERTICAL MENU key.
(15) Press main-menu Fine Scale key.
(16) Use front-panel keypad to enter 1 mV vertical Fine Scale setting. Press
front panel 1 key.
(17) Press front panel SHIFT key.
(18) Press front panel m key.
(19) Press front panel ENTER key.
(20) Press front-panel CLEAR MENU key.
CHANGE 1 11