TB 9-6625-2320-35
(26) Adjust front-panel HORIZONTAL SCALE knob to display a delay (D) time
base of 500 s.
(27) Press front-panel SHIFT key.
(28) Press front-panel DELAYED TRIG key.
(29) Press main-menu Level key.
NOTE
Notice in the following step the side-menu SET TO 50% key
is called for, not the front-panel SET LEVEL TO 50% key.
(30) Press side-menu SET TO 50% key.
(31) If TI main-menu Level readout does not indicate between 9.9393 and 10.1147
V, perform b below.
(32) Press main-menu Slope key.
(33) Press side-menu negative slope
key.
(34) Press main-menu Level key.
NOTE
Notice in the following step the side-menu SET TO 50% key
is called for, not the front-panel SET LEVEL TO 50% key.
(35) Press side-menu SET TO 50% key.
(36) If TI main-menu Level readout does not indicate between 9.9393 and 10.1147
V, perform b below.
16. Sensitivity and Edge Trigger
a. Performance Check
(1) Press front-panel SETUP key.
(2) Press main-menu Recall Factory Setup key.
(3) Press side-menu OK Confirm Factory Init key.