TB 9-6625-2320-35
18. Probe Compensator Output
a. Performance Check
(1) Connect equipment as shown in figure 8 below.
(2) Press front-panel SETUP key.
(3) Press main-menu Recall Factory Setup key.
(4) Press side-menu OK Confirm Factory Init key.
(5) Adjust front-panel HORIZONTAL SCALE knob to display 200 s.
(6) Press front-panel SET LEVEL TO 50% key.
(7) Adjust front-panel VERTICAL POSITION knob to center test signal.
(8) Press front-panel SHIFT key.
(9) Press front-panel ACQUIRE MENU key.
(10) Press main-menu Mode key.
(11) Press side-menu Average key.
(12) Adjust general purpose knob to set averages to 128.
(13) Press front-panel MEASURE key.