TB 9-6625-2320-35
(29) Position calibrator for a 0 V output.
(30) Press front-panel Shift key.
(31) Press front-panel ACQUIRE MENU key.
(32) Press main-menu Mode key.
(33) Press side-menu Average key.
(34) Adjust general purpose knob to set averages to 16.
(upper) level of stored compensation signal. Value will be near 0.5 V. Record value.
(36) DELETE
(lower) level of stored compensation signal. Value will be near 0 V. Record value.
(38) Press front-panel CLEAR MENU key.
(39) Subtract the lower level value from the top level value. If value does not
indicate between 495 and 505 mV, perform b below.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750- 25.
ADJUSTMENT PROCESS
20. Preliminary Instructions. The
procedure in
paragraphs 21 through
22 should be
performed only if an out-of-tolerance condition exists in paragraphs 8 through 19 above.
WARNING
HIGH VOLTAGE is used or exposed during the performance
of this calibration. DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Connect TI to a 115 V ac source.
PIN: 076015-001