TB 9-6625-2320-35
d. When "Installation Complete" appears, remove installation disk and store in a
secure place.
NOTE
Use a editor of your choice to make the following changes to
the ADJ700.BAT file located in the TDS700.ADJ directory.
e. Change directories to the
TDS700.adj directory and edit the
following lines in the
ADJ700.BAT file:
(1) Change SET GBIP0=PC2 0 0 68 to SET GBIP0=. (Delete PC2 0 0 68)
(2) Change SET GBIP1= to SET GBIP1=PC2 0 0 68. (Add PC2 0 0 68)
f. Save modified ADJ700.BAT file and reset the controller.
NOTE
If the TDS 500B/700A Field Adjustment Software is already
installed on the PC, go to the TDS700.ADJ drive and
directory where the software is located.
g. At the TDS700.ADJ directory prompt, type ADJ700 to initiate the adjustment
process.
CAUTION
The following adjustment process must be completed in it's
entirety before deenergizing TI. If the adjustment process
cannot be completed, keep TI energized until adjustment
process can be completed.
h. Enter 7 to select the TDS540B.
i. Enter 1 to select RUN FULL SEQUENCE.
NOTE
Refer to the following diagram for the MEMORY WRITE
PROTECT SWITCH referenced in the adjustment process.
38