TB 9-6625-2323-35
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual and ST400-AW-MMC-
020 and ST400-HW-MMC-030 for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(s) to minimum after each step within the
performance check where applicable.
a. Remove TI protective cover only as necessary to make adjustments.
NOTE
For the remainder of this procedure the 68369NV connected to
the signal generator workstation will be called the local
CAUTION
Before connecting TI to power source, make sure TI is set to the
power source line voltage as shown on rear of TI.
b. Connect TI to 115 V ac power source.
c. Press LINE switch to OPERATE and allow at least one hour for TI to warm-up and stabilize.