TB 9-6625-2323-35
(14) Reduce all outputs to minimum.
(15) Disconnect equipment setup.
b. Adjustments. No further adjustments can be made.
a. Performance Check
(1) Connect equipment as shown in figure 5 below.
(2) Set measuring receiver to measure frequency modulation, with +PEAK detector,
<20 Hz high
pass and
>200 kHz Lo
pass filters and at a
frequency of 5
GHz.
(3) Press TI keys as listed in (a) through (j) below.
(a)
System, Reset.
(b)
Output On Off to off.
(c)
Level Control, L1, Edit L1, 0, and MHz ms dB.
(d)
Frequency Control, F0, Edit F0, 5, and GHz Sec dBm.
(e)
Output On Off to on.
(f)
Modulation, FM, More, FM Modes, Locked, Previous Menu, Previous Menu.
(g)
Edit Dev., 2, 4, 0, kHz s Step, Edit Dev.
Mod. Wave, arrow key to ~, Select, Previous Menu.
(h)
(i)
Edit Rate, 1, 0, 0, kHz s Step, Edit Rate.
(j)
On/Off to on.