TB 9-6625-2328-24
b. Adjustments
(1) Disconnect pulse generator from TI.
(2) Press and release TI POWER ON/STBY pushbutton to STBY.
(3) Press and hold MATH Scale & Offset key while pressing POWER ON/STBY
pushbutton to ON until CAL SECURE is displayed.
(4) Press MATH Scale & Offset key until CODE:
0 is displayed.
(5) Press ⇐ key until CODE: 00000 is displayed.
(6) Press , , ⇐, keys until CODE: 53131 is displayed then press Enter key
and verify CAL UNSECURE is displayed.
Function
Outputs
Unbalanced
(7) Connect function
generator
to
TI
CHANNEL 1 input.
(8) Set function generator for a square wave, 10 MHz, 1 V p-p, into a 50 Ω feed
through.
(9) Press MATH Scale & Offset key until CAL:OFFS 1 ? is displayed.
(10)Press key until CAL: TI QUIK ? is displayed.
(11)Press Enter key. When THEN ENTER is displayed, press Enter key.
(12) Verify CALIBRATING is displayed followed by TI CAL PASS.
(13)Press MATH Scale & Offset key until CODE:
0 is displayed.
(14)Press ⇐ key until CODE: 00000 is displayed.
(15)Press , , ⇐, keys until CODE: 53131 is displayed then press Enter key
and verify CAL SECURE is displayed.
(16)Disconnect function generator from TI.
a. Performance Check
(1) Press and release TI POWER ON/STBY pushbutton to STBY.
(2) Press TI POWER ON/STBY pushbutton to ON.
(3) Connect function generator Output to TI CHANNEL 1 input.
(4) Press TI keys as listed in (a) through (h) below:
(a) CHANNEL 1 Trigger/Sensitivity until AUTO TRG: is displayed, then
to display AUTO TRG: OFF.
(b) CHANNEL 1 Trigger/Sensitivity until LEVEL: is displayed, if necessary,
use , , ⇐, to display LEVEL: 0.000 V and then Enter.
(c) CHANNEL 1 50Ω/1MΩ to 50Ω (LED on).
(d) CHANNEL 1 DC/AC to DC (LED on).
(e) CHANNEL 2 Trigger/Sensitivity until AUTO TRG: is displayed, then
to display AUTO TRG: OFF.
(f) CHANNEL 2 Trigger/Sensitivity until LEVEL: is displayed, if necessary,
use , , ⇐, to display LEVEL: 0.000 V and then Enter.