TB 9-6625-2328-24
a. Performance Check
(1) Press and release TI POWER ON/STBY pushbutton to STBY.
(2) Press TI POWER ON/STBY pushbutton to ON.
(3) Connect calibrator OUTPUT HI and LO to TI CHANNEL 1 input.
(4) Press TI keys as listed in (a) through (c) below:
(a) MEASURE Other Meas until VOLT PEAKS 1 is displayed.
(b) CHANNEL 1 DC/AC to DC (LED on).
(c) CHANNEL 1 100kHz Filter to on (LED on).
(5) Set calibrator for a 2 V dc output. TI will indicate +2.00 V 0.06 V and +2.00 V 0.06 V.
(6) Set calibrator for a -2 V dc output. TI will indicate -2.00 V 0.06 V and -2.00 V 0.06 V.
(7) Press TI CHANNEL 1 DC/AC key to select AC (LED off) and CHANNEL 1
100kHz Filter key to off (LED off).
(8) Set calibrator for a 0.707 V, 1 MHz output. TI will indicate 1.00 V 0.12 V and
+1.00 V 0.12 V.
(9) Connect calibrator OUTPUT HI and LO to TI CHANNEL 2 input.
(10)Press TI keys as listed in (a) through (c) below:
(a) MEASURE Other Meas until VOLT PEAKS 2 is displayed.
(b) CHANNEL 2 DC/AC to DC (LED on).
(c) CHANNEL 2 100kHz Filter to on (LED on).
(11) Set calibrator for a 2 V dc output. TI will indicate +2.00 V 0.06 V and +2.00 V
0.06 V.
(12) Set calibrator for a -2 V dc output. TI will indicate -2.00 V 0.06 V and -2.00 V
0.06 V.
(13) Press TI CHANNEL 2 DC/AC key to select AC (LED off) and CHANNEL 2
100kHz Filter key to off (LED off).
(14) Set calibrator for a 0.707 V, 1 MHz output. TI will indicate 1.00 V 0.12 V and
+1.00 V 0.12 V.
(15) Disconnect calibrator from TI.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Press and release TI POWER ON/STBY pushbutton to STBY.
(2) Press TI POWER ON/STBY pushbutton to ON.
(3) Connect time/frequency workstation OUTPUT 10 MHz to TI CHANNEL 1 input.
(4) Press TI keys as listed in (a) through (f) below: