TB 9-6625-2328-35
(8) Set function generator for a square wave, 10 MHz, 1 V p-p, 50Ω output.
(9) Press MATH Scale & Offset key until CAL:OFFS 1 ? is displayed.
(10) Press ⇒ key until CAL: TI QUIK ? is displayed.
(11) Press Enter key. When THEN ENTER is displayed, press Enter key.
(12) Verify CALIBRATING is displayed followed by TI CAL PASS.
(13) Press MATH Scale & Offset key until CODE:
0 is displayed.
(14) Press ⇐ key until CODE: 00000 is displayed.
(15) Press ⇒ , ⇓ , ⇐ , ⇑ keys until CODE: 53131 is displayed then press Enter key
and verify CAL SECURE is displayed.
(16) Disconnect function generator from TI.
a. Performance Check
(1) Press and release TI POWER ON/STBY pushbutton to STBY.
(2) Press TI POWER ON/STBY pushbutton to ON.
(3) Connect function
generator
Function
Outputs
Unbalanced to
TI
CHANNEL 1 input.
(4) Press TI keys as listed in (a) through (h) below:
(a) CHANNEL 1 Trigger/Sensitivity until AUTO TRG: is displayed, then ⇒
to display AUTO TRG: OFF.
(b) CHANNEL 1 Trigger/Sensitivity until LEVEL: is displayed, if
necessary, use ⇒ , ⇑ , ⇐ , ⇓ to display LEVEL: 0.000 V and then Enter.
(c) CHANNEL 1 50Ω /1MΩ to 50Ω (LED on).
(d) CHANNEL 1 DC/AC to DC (LED on).
(e) CHANNEL 2 Trigger/Sensitivity until AUTO TRG: is displayed, then ⇒
to display AUTO TRG: OFF.
(f) CHANNEL 2 Trigger/Sensitivity until LEVEL: is displayed, if
necessary, use ⇒ , ⇑ , ⇐ , ⇓ to display LEVEL: 0.000 V and then Enter.
(g) CHANNEL 2 50Ω /1MΩ to 50Ω (LED on).
(h) CHANNEL 2 DC/AC to DC (LED on).
(5) Set function generator for a square wave, 1 MHz, 80 mV p-p, -60 mV dc offset,
50Ω output.
(6) Press TI RUN key.