TB 9-6625-2328-35
(4) Press TI keys as listed in (a) through (j) below:
(a) MEASURE Gate & ExtArm until GATE: is displayed; if necessary, press
⇒ until GATE: TIME is displayed.
(b) MEASURE Gate & ExtArm until TIME: is displayed; use ⇒ , ⇑ , ⇐ , ⇓ to
display TIME: 1.000 s and then Enter.
(c) CHANNEL 1 Trigger/Sensitivity until AUTO TRG: is displayed, then ⇒
to display AUTO TRG: OFF.
(d) CHANNEL 1 Trigger/Sensitivity until LEVEL: is displayed, if
necessary, use ⇒ , ⇑ , ⇐ , ⇓ to display LEVEL: 0.000 V and then Enter.
(e) CHANNEL 1 50Ω /1MΩ to 50Ω (LED on).
(f) CHANNEL 1 DC/AC to DC (LED on).
(g) CHANNEL 2 Trigger/Sensitivity until AUTO TRG: is displayed, then ⇒
to display AUTO TRG: OFF.
(h) CHANNEL 2 Trigger/Sensitivity until LEVEL: is displayed; if
necessary, use ⇒ , ⇑ , ⇐ , ⇓ to display LEVEL: 0.000 V and then Enter.
(i) CHANNEL 2 50Ω /1MΩ to 50Ω (LED on).
(j) CHANNEL 2 DC/AC to DC (LED on).
(5) Connect time/frequency workstation OUTPUT 1 MHz to TI CHANNEL 1
input.
(6) Press TI Stop/Single key. TI will indicate between 999,999.999 and
1,000,000.001 Hz.
(7) Disconnect time/frequency workstation OUTPUT 1 MHz from TI CHANNEL
1 input.
(8) Connect signal generator OUTPUT RF to TI CHANNEL 1 input.
(9) Set signal generator for a 100 MHz, -7 dBm output.
11