TB 9-6625-2328-35
(10) Press TI Stop/Single key. TI will indicate between 99.999,999,97 and
100.000,000,03 MHz.
(11) Set signal generator for a 200 MHz, -7 dBm output.
(12) Press TI Stop/Single key. TI will indicate between 199.999,999,94 and
200.000,000,06 MHz.
(13) Set signal generator for a 225 MHz, -7 dBm output.
(14) Press TI Stop/Single key. TI will indicate between 224.999,999,93 and
225.000,000,07 MHz.
(15) Disconnect signal generator OUTPUT RF from TI CHANNEL 1 input.
(16) Connect time/frequency workstation OUTPUT 1 MHz to TI CHANNEL 2
input.
(17) Press TI MEASURE Freq & Ratio key until TI displays FREQUENCY 2.
Repeat (6) through (15) above for CHANNEL 2.
(18) Connect signal generator OUTPUT RF to TI CHANNEL 3 input.
(19) Press TI MEASURE Freq & Ratio key until FREQUENCY 3 is displayed.
(20) Set signal generator for a 100 MHz, -17 dBm output.
(21) Press TI Stop/Single key. TI will indicate between 99.999,999,97 and
100.000,000,03 MHz.
(22) Set signal generator for a 1300 MHz, -17 dBm output.
(23) Press TI Stop/Single key. TI will indicate between 1,299.999,999,5 and
1,300.000,000,5 MHz.
(24) Disconnect equipment from TI.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Press and release TI POWER ON/STBY pushbutton to STBY.
(2) Press TI POWER ON/STBY pushbutton to ON.
(3) Connect calibrator OUTPUT HI and LO to TI CHANNEL 1 input.
(4) Press TI keys as listed in (a) through (c) below:
(a) MEASURE Other Meas until VOLT PEAKS 1 is displayed.
(b) CHANNEL 1 DC/AC to DC (LED on).
(c) CHANNEL 1 100kHz Filter to on (LED on).
(5) Set calibrator for a 2 V dc output. TI will indicate +2.00 V 0.06 V and +2.00 V
0.06 V.