TB 9-6625-2329-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Signal Generator, Rhode And Schwarz, Model 801.0001.43.
The
manufacturer's manual was used as the prime data source in compiling these instructions.
The equipment being calibrated will be referred to as the TI (test instrument) throughout
this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 6
hours, using the dc and low frequency and microwave technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Range: 100 kHz to 1000 MHz
Accuracy: RF > 31.25 MHz (+0.5 x 10-9) + ref error
Accuracy: RF < 31.25 MHz (+0.1 Hz + ref error)
Time base stability: +1 x 10-9 per day
RF output
Range: 13 dBm
Accuracy: +1.5 dB
Flatness range: 100 kHz to 1000 MHz
Accuracy: <1 dB at 0 dBm
Range: 0 to 120 dB
Accuracy: +1.5 dB
Range: 0 to 20 dB
Accuracy: +0.2 dB at 5 dB
+0.5 dB at 10 dB
+0.5 dB at 15 dB
+0.5 dB at 20 dB
Spectral purity
Range: 100 kHz to 1000 MHz
Accuracy: <-30 dBc
Spurious signal range:
Accuracy
100 kHz to 31.25 MHz
<-70 dBc
31.25 MHz to 250 MHz
<-80 dBc
250 MHz to 1000 MHz
<-70 dBc
See footnotes at end of document.